Piezoelectric accelerometer comparison calibration can be divided into simple comparison back-to-back calibration, using FFT analysis method of comparing back-to-back calibration, FFT is back-to-back calibration three alternative method.
Computer program_controlled automatic accelerometer vibration calibration system with computer program and FFT analysis technology, precise measurement by the school and the known accelerometer, the transfer function between the accelerometer dynamic calibration, replacement method is used to improve FFT calibration at the same time, improve the accuracy of measurement comparison calibration. Uncertainty of measurement in 20 ~ 2000 hz is 1.5%, overcome the single frequency of simple comparison back-to-back calibration calibration and FFT analysis method to compare back-to-back calibration measurement uncertainty of big shortcoming.
Alternative measure calibration method needs two back_to_back, first measured and stored work standard accelerometer and reference standard accelerometer the transfer function between the Hr (f), and then measure and store the tested accelerometer and work standard accelerometer transfer function between Hu (f), the technology will offset the most caused by vibration table distortion tolerance and measuring instrument of uncertainty components. In two measurement, work standard accelerometer installation on the vibration table all the time, while the standard reference accelerometer and the tested accelerometer, respectively, with the work standard accelerometer is used in the comparison. By testing the sensitivity of the accelerometer Su = Sr (f) (f) [Hu (f)/Hr (f)], type, Su (f) to test the sensitivity of the accelerometer; The Sr (f) as the reference standard, the sensitivity of the accelerometer; Hu (f) for the tested accelerometer and work standard accelerometer frequency transfer function; Hr (f) as the standard reference accelerometer and work standard frequency transfer function between the accelerometer. Ratio of Hu (f)/Hr (f) given by the frequency response function of compensation. With plural than f calculation.
Eliminate the error caused by electronic instruments with alternative technology, improve the calibration accuracy of back_to_back in a broad band range, measurement uncertainty is better than 1.5%. Alternative method of calibration for reference standard accelerometer calibration, reference standard accelerometer as roots and transmission, and not as a working standard used in the comparison calibration of accelerometer and accelerometer being measured directly in the standard reference accelerometer, but this is to reduce the damage probability of standard reference accelerometer.
Using standard accelerometer, 8305 (748321) on the vibration table with a simple calibration method and back_to_back on the 28978 automatic accelerometer calibration system with alternative method of FFT calibration frequency response test, test values and the verification department with the method of absolute calibration of the measurement result comparison, FFT calibration of accelerometer with alternative method in 20 ~ 2000 hz within the scope of the calibration error is 1.2%, and can be calibrated to 10 KHZ; And simple back_to_back calibration error of 2.2%. Therefore, 28978 accelerometer calibration by the alternative method of vibration calibration system work for calibration of accelerometer provides a more simple, more reliable calibration method.